Instalab® 600 Product Support
Below are the support materials available for the Instalab® 600.
A: An error 6 means system or excessive reference log drift. If the problem does not clear on its own, clean the reference disk and aperture glass.
A. An error 7 means there is too much system noise. It will normally clear itself, but if it doesn't, contact DICKEY-john service at 1-800-637-3302.
A. Yes, DICKEY-john has a regression program that will help you develop your own constants.
A. No, it can only test ground samples.
A: Error 0 means the sample is not reflective enough. Clean the sample cup, or try a different sample.
A: Error 11 means a constant has been altered in memory. Check to see if a constant was changed.
A. Typically, the INSTALAB will test a sample in 10 seconds.
A. Wait a couple of hours for the machine to warm up.
A. Error 1 means the sample is too reflective. Clean the sample cup, or try a different sample.
A. The INSTALAB can use 4-10 filters.
- Instalab® 600 literature
- Error Codes
- Instalab Regression and Data Collection Program Manual
- Instalab® Operator's Manual
- Instalab Regression software